The GloveFAST Cyto Pharma uses a closed architecture with active pressure control and vertical laminar flow. The work chamber is maintained at -75 Pa and the transfer airlocks at -110 Pa. Airlock doors are interlocked to prevent simultaneous opening and reduce cross-contamination risk.
Configuration: two transfer airlocks, 4 gloves
External dimensions: 2815 × 2350 × 868 mm (W × H × D)
Usable dimensions: 1497 × 740 × 580 mm (W × H × D)
Internal airlock-door dimensions: 370 × 338 mm
External airlock-door dimensions: 370 × 324 mm
Glove-port diameter: 300 mm
Work-chamber pressure: -75 Pa
Airlock pressure: -110 Pa
Laminar airflow velocity: 0.45 m/s
Exhaust air volume: 200 m³/h
Noise level: ≤ 54 dB(A)
Light intensity: > 2,000 lux
Connected load: ≤ 280 W
Weight: 480 kg
Construction: Epoxy-Painted Steel
Power supply: 230 V, 50 Hz
Key features
Cytotoxic containment
Negative pressure and a closed system for cytotoxic, CMR and highly potent substances.
Two transfer airlocks
Interlocked doors, pressure control and a high air-exchange rate for safe material transfer.
4 glove positions
Complete physical separation between operator and process, with 300 mm glove ports.
H14 HEPA filtration
Vertical laminar flow and continuous monitoring of supply and exhaust airflow.
Epoxy-Painted Steel construction
Electropolished AISI 316L stainless-steel work surface, durable and easy to decontaminate.
GMP compliance
HMI panel, pressure-hold testing, IQ/OQ documentation and options for 21 CFR Part 11 and H₂O₂ decontamination.